Click here to view the clickable sidebar

The IAC Standards and Guidelines
for Nuclear/PET Accreditation

 

Click here for a printer-friendly PDF of the Nuclear/PET Standards

Part B:
Examinations and Procedures

Section 1B: Instrumentation and Equipment

 

STANDARD – Instrumentation

 

1.1B           Equipment and instrumentation used in the nuclear medicine facility must be in good working condition and must be routinely inspected for safety and proper functionality and records kept on file.

 

1.1.1B            All imaging and non-imaging devices must be FDA-approved or used under an approved research protocol with informed consent by the patient.

 

1.1.2B            The facility must maintain records of service and maintenance.

 

1.1.3B            Equipment and instrumentation must include at least the following:

 

1.1.3.1B             dose calibrator or decay correction calculation system, as applicable;

 

1.1.3.2B             imaging/counting equipment;

 

1.1.3.3B             radiation monitoring devices including:

 

i.               portable survey meter (required);

ii.             removable contamination counting equipment (as applicable);

iii.           fixed area survey meter for dose preparation/storage areas (as applicable).

 

1.1.3.4B             resuscitation equipment and supplies (appropriate to the types of procedures being performed):

 

i.               oxygen;

ii.             defibrillator/AED;

iii.           emergency drugs (including a master list; all unexpired).

 

1.1.3.5B             exercise equipment (as applicable);

 

1.1.3.6B             ECG equipment (as applicable);

 

1.1.3.7B             ancillary monitoring equipment (as applicable);

 

1.1.3.8B             infusion pumps/automated injectors (as applicable);

 

1.1.3.9B             glucometers (as applicable);

 

1.1.3.10B          hood for volatile radionuclides or cell handling (as applicable);

 

1.1.3.11B          xenon (or other gas) trap (as applicable).


1.1.3.12B          Rubidium Rb82 generator and direct infusion system (as applicable).

 

STANDARD – Equipment Quality Control Protocols

 

1.2B           The facility must have acceptable site-specific written protocols for all routine quality control procedures of imaging and non-imaging equipment.

 

Comment: Simply stating “following manufacturer’s recommendations” is not sufficient.

 

1.2.1B            The facility must maintain records of all routine quality control of imaging and non-imaging equipment.

 

1.2.1.1B             Protocols for quality control of imaging equipment must be step-by-step, camera-specific and include:

 

i.               frequency of test performance;

ii.             type of source and position (e.g., point, sheet), if applicable;

iii.           equipment setup (e.g., intrinsic, extrinsic, collimator, energy window setting, matrix size, zoom, etc.);

iv.           acquisition instructions including views, time/counts, etc.;

v.             processing instructions, if applicable (e.g., region of interest for quantification, sinogram generation, graphs, etc.);

vi.            evaluation and acceptable range (or tolerance limits) of the results of each procedure;

vii.          instructions for corrective action of out-of-tolerance results;

viii.        instructions for retention and comparison with previous results.

 


1.2.1.2B             Protocols  for quality control of non-imaging equipment must be step-by-step and include:


i.               frequency of testing performance;

ii.             sources to be used, if applicable;

iii.           data to be collected with acceptable ranges and timing required;

iv.          instructions for corrective action of out-of-tolerance results.



1.2.1.3B             The results of QC testing must be reviewed by appropriate staff in a timely manner and action taken if results are not within tolerance limits.

 

1.2.1.4B               Quality control protocols must be reviewed and/or updated as equipment is changed and at least every three years by the Medical Director, physicist or other responsible person.

 

1.2.2B            If frequency of QC testing varies from the above, justification must be based on scientific data or manufacturer’s recommendation. If a less frequent schedule is being used, there must be clear documentation of the justification (such as based on scientific data).

 

1.2.3B            Appropriate reference standards (i.e., sealed sources) for QC of imaging and non-imaging equipment must be used with a reference source traceable to the National Institute of Standards and Technology (NIST).

 

(See Guidelines below for further recommendations.)

STANDARD – Imaging Equipment Quality Control

 

1.3B            Site-specific, detailed protocols must be documented and followed for routine inspection and testing of all imaging equipment. Protocols must be in accordance with all applicable federal, state and local requirements.

 

1.3.1B            Gamma Camera (Planar, SPECT, and SPECT/CT)

1.3.1.1B             Energy peaking to verify that the photopeak is centered in the set photopeak energy window must be performed, if applicable (documentation not required).
Frequency: Daily (prior to use) or per manufacturer’s recommendation

1.3.1.2B             Intrinsic or extrinsic uniformity calculation of integral and/or differential uniformity value must be performed (e.g., 3-5%).
Frequency: Daily (prior to use)

1.3.1.3B             Spatial resolution/spatial linearity with resolution phantom (e.g., bars) must be performed.
Frequency: Weekly

 

1.3.1.4B             Center-of-rotation (COR) must be performed on SPECT cameras to ensure mechanical and electrical alignment of the center of field of view.
Frequency: Monthly

 

1.3.1.5B             High-count flood for uniformity correction, performed to correct for residual detector and collimator non-uniformity, must be performed.
Frequency: Per manufacturer’s recommendation

 

1.3.1.6B             Preventive maintenance (PM) must be performed.
Frequency: Every six months

 

1.3.1.7B               For SPECT/CT equipment, daily system tests (as recommended by the manufacturer) must be performed to assess system function/constancy (e.g., calibration scan, CT warm-up, CT calibration, water phantom, coincidence timing, normalization update, etc.). The daily system test procedure must be specifically described.

 

Comment: Energy peaking and uniformity testing must be appropriate for the energy of the radioisotopes being imaged (e.g., low energy or medium energy).

 

Comment: If imaging equipment is physically moved from site to site, (other than planar mobile gamma cameras or non-PMT mobile planar/SPECT cameras used within a building) the QC tests must be repeated after each move and prior to equipment use.

 

(See Guidelines below for further recommendations.)

 

 

1.3.2B            PET and PET/CT Scanner

 

1.3.2.1B             Daily system tests (as recommended by the manufacturer) must be performed to assess system function/constancy (e.g., calibration scan, blank scan, CT warm-up, CT calibration, water phantom, coincidence timing, normalization update, etc.). The daily system test procedure must be specifically described.
Frequency: Daily (prior to use)

 

1.3.2.2B             Tomographic uniformity using a cylinder phantom of uniform radioactivity must be performed.
Frequency: Per manufacturer’s recommendation

 

1.3.2.3B             Normalization to calibrate the efficiency of all detectors in the system must be performed.
Frequency: Per manufacturer’s recommendation and after preventive maintenance and major hardware repair.

 

1.3.2.4B              For facilities performing quantitative analysis of images, such as standardized uptake value (SUV) calculation, absolute activity calibration and testing must be performed. Facilities must develop procedures that validate quantitative scanner performance.
Frequency: After a hardware change or per manufacturer’s recommendations

 

1.3.2.5B             Preventive maintenance of all PET and PET/CT scanners must be performed. Frequency: Every six months

 

Comment: If the PET or PET/CT scanner is physically moved from site to site, the daily system test as described in Standard 1.3.2.1B must be performed after each scanner relocation and prior to injection.

 

(See Guidelines below for further recommendations.)

 

1.3.3B            CT-Specific Quality Control - Please refer to the IAC Standards and Guidelines for CT Accreditation for specific quality control requirements for CT scanners.

STANDARD – Non-imaging Equipment Quality Control

 

1.4B           Site-specific, detailed protocols must be documented and followed for routine inspection and testing of all non-imaging equipment. Protocols must be in accordance with all applicable federal, state and local requirements.

 

1.4.1B            Survey Meter

 

1.4.1.1B             Constancy of response must be checked by measuring the exposure or counting rate of a long-lived reference source. Measurements must be within acceptable tolerance levels (e.g., within 10-20%).
Frequency: Daily or prior to use or per manufacturer’s recommendation

 

1.4.1.2B             The battery must be checked, if applicable, to verify the voltage supplied by the battery is within the acceptable operating range.
Frequency: Daily or prior to use

 

1.4.1.3B             The survey meter must be calibrated using suitable long-lived reference sources. Frequency: Annual or following repair as per manufacturer’s recommendation

(
See Guidelines on Page 40 for further recommendations.)

 

1.4.2B            Dose Calibrator

 

1.4.2.1B             Background exposure must be measured
Frequency: Daily or before use

Comment: This must include daily quality control as required by the manufacturer’s recommendations.

 

1.4.2.2B             Constancy of response must be checked by measuring the exposure or counting rate of a long-lived reference source. Measurements must be within acceptable tolerance levels (e.g., within 5-10% or per manufacturer’s recommendation).
Frequency: Daily or before use

 

1.4.2.3B             Linearity that is within tolerance limits must be verified (e.g., within 10%). Method of linearity check (i.e., decay or shield method) including activity, volume, time of measurement, etc., must be specifically defined.
Frequency: Quarterly

 

1.4.2.4B             Accuracy that is within tolerance limits must be verified (e.g., within 5-10% or per manufacturer’s recommendation).
Frequency: Annual

(See Guidelines below for further recommendations.)

 

1.4.2.5B             Measurement of geometry dependent responses that affect volume changes of vials and/or syringes must be performed.
Frequency: At installation and following repair or relocation as per manufacturer’s recommendation.

 

 

 

1.4.3B            Well Counter

1.4.3.1B             Energy spectrum check, if applicable, to verify that the counter is properly peaked and that the photopeaks of the radionuclides coincide with the preset photopeak energy windows.

Frequency: Daily or per manufacturer recommendation

1.4.3.2B             Background exposure or counting rate must be measured.
Frequency: Daily (or prior to use)

 

1.4.3.3B             Constancy of response must be checked by measuring the exposure or counting rate of a long-lived reference source. Measurements must be within acceptable tolerance levels (e.g., within 5-10% as per manufacturer’s recommendation).
Frequency: Daily (or prior to use)

 

1.4.3.4B             Chi-square (X2) test, if applicable, to measure reproducibility and random variation must be performed.
Frequency: Quarterly

 

1.4.3.5B             Efficiency to determine the ratio of detected counts measured by the system to the actual rate of decay (cpm/mCi or dpm), for a specific nuclide or region of interest must be performed.
Frequency: Annual or per manufacturer’s recommendation

 

1.4.4B            Intraoperative Probes

 

1.4.4.1B             The battery must be checked to verify the voltage supplied by the battery is within the acceptable operating range.
Frequency: Daily or before use

 

1.4.4.2B             Background exposure or counting rate must be measured.
Frequency: Daily or before use

 

1.4.4.3B             Bias voltage of primary and back-up battery must be checked, if applicable. Frequency: Per manufacturer’s recommendation

 

1.4.4.4B             Constancy of response must be checked by measuring the exposure or counting rate of a long-lived reference source. Measurements must be within acceptable tolerance levels (e.g., within 5-10% or as per manufacturer’s recommendation).
Frequency: Daily or before use

 

1.4.5B            Organ Uptake Probes (e.g., thyroid uptake probes)

 

1.4.5.1B             System Test/Detector Status/Autocalibration (as recommended by manufacturer) must be performed to assess internal data, full width half-maximum (FWHM), voltage and gain settings.
Frequency: Daily or before use

 

1.4.5.2B             Energy spectrum check, if applicable, to verify that the counter is properly peaked and that the photopeaks of the radionuclides coincide with the preset photopeak energy windows.
Frequency: Daily or before use

 

1.4.5.3B             Background exposure or counting rate must be measured.
Frequency: Daily or before use

1.4.5.4B             Constancy of response must be checked by measuring the exposure or counting rate of a long-lived reference source. Measurements must be within acceptable tolerance levels (e.g., within 5-10% or as per manufacturer’s recommendation).
Frequency: Daily or before use

1.4.5.5B             If probe is used to perform radioactive contamination wipe tests, efficiency for a specific nuclide or region of interest must be measured to determine the ratio of detected counts measured by the system to the actual rate of decay (cpm/Bq or cpm/mCi) or disintegrations per minute (dpm).
Frequency: Annual or per manufacturer’s recommendation

 

1.4.5.6B             Chi-square (X2) test to measure reproducibility and random variation must be performed.
Frequency: Quarterly


1.4.6B            Organ Uptake Probes (e.g., thyroid uptake probes)

 

1.4.6.1B             System Flush
Frequency: Daily or before use

 

1.4.6.2B             Dose Calibrator Calibration
Frequency: Daily or before use

 

1.4.6.3B             Sr breakthrough testing
Frequency: Daily or before use


1.4.6.B             Generator Replacement / Installation
Frequency: Per viable life of generator

STANDARD – Other Equipment Quality Control

1.5B           Site-specific, detailed protocols must be documented and be followed for routine inspection and testing of all other medical equipment. Protocols must be in accordance with all federal, state and local requirements.

 

1.5.1B            Emergency Equipment

 

1.5.1.1B             An emergency response cart or kit, appropriate for the types of procedures being performed, must be present. There must be documentation that it is checked to assure that all expected items are present and none is expired.
Frequency: Monthly

 

1.5.1.2B             Defibrillator/AED device and supplies (e.g., pads, gel) must be checked for functionality (e.g., voltage and battery, expiration date)
Frequency: Daily when patient studies are performed

 

1.5.1.3B             Oxygen sources (wall unit or portable cylinder) must be checked for availability, pressure gauge shows adequate tank filling, proper function and proper tubing/mask.
Frequency: Daily when patient studies are performed

 

1.5.2B            Miscellaneous Equipment

 

1.5.2.1B             Glucometer accuracy must be confirmed.
Frequency: Daily if used

 

1.5.2.2B             Infusion pump accuracy must be confirmed.
Frequency: Per manufacturer’s recommendation

 

1.5.2.3B             Xenon trap and nebulizer

 

i.               Nebulizer must be visually inspected for damage and cleaned.
Frequency: As necessary

ii.             Xenon trap moisture absorbing crystals must be replaced (e.g., every 3-5 patients).
Frequency: Per manufacturer's recommendation

iii.           Xenon trap leak test must be performed.
Frequency: Monthly

iv.           Xenon trap/charcoal filters must be replaced.
Frequency: Per manufacturer’s recommendation

 

 

 

Section 1B: Instrumentation and Equipment
Guidelines

 

1.2B            For each quality control test performed, the following information should be recorded:

 

·         the test performed;

·         date and time of the test;

·         identification of the device tested (e.g., make, model);

·         the make, model and serial number of any reference sources used, if applicable;

·         the results of the test;

·         a notation indicating if the test result was or was not acceptable;

·         the signature or initials of the individual performing the test or clear delineation of this duty written into a policy.

 

Comment: Preferably, the information is recorded on a structured form or documented in a facility management program.

 

Initial acceptance results for all equipment should be retained and used for comparison. Preferably, acceptance testing should be performed by a party other than the equipment supplier

 

1.3.1B        Gamma Camera:

 

Overall system performance may be evaluated using a fillable phantom containing non-radioactive (cold) inserts of different sizes and visually inspecting the resulting images.
Frequency: Annually

 

Collimator integrity, comparing the extrinsic and intrinsic uniformity flood along with visual inspection of collimator for damage, should be performed.
Frequency: Annually

 

1.3.2B        PET and PET/CT Scanners:

 

Bed position overlap (e.g., 2-bed test, continuous bed motion, incremental bed overlap) is recommended. Frequency: Per manufacturer recommendation

 

Alignment of the PET and CT scans (for PET/CT scanners) should be performed as per manufacturer recommendation.
Frequency: Per manufacturer recommendation

 

1.4.1.3B     Survey Meter - A dated sticker summarizing the calibration results should be affixed to the meter itself. The calibration report should specify the reference sources, the measurement procedure and the measured and expected exposure rates.

 

1.4.2.4B     Dose Calibrator - It is preferable that accuracy be measured with at least two reference sources.