Click here to view the clickable sidebar

The IAC Standards and Guidelines
for Adult Echocardiography Accreditation

 

Click here for a printer-friendly PDF of the Adult Echocardiography Standards

Part B:
Examinations and Procedures

Section 1B: Adult Transthoracic Echocardiography Testing

 

STANDARD – Instrumentation

 

1.1B         Cardiac Ultrasound Systems

 

1.1.1B         Ultrasound instruments utilized for diagnostic studies must include, at a minimum, hardware and software to perform:

 

1.1.1.1B        M-Mode imaging

 

1.1.1.2B        2-D imaging. The system must include harmonic capabilities.

 

1.1.1.3B        Spectral display for pulsed (PW) and continuous wave (CW) Doppler and tissue Doppler imaging;

 

1.1.1.4B       color flow Doppler;

 

1.1.1.5B        Monitor or other display method of suitable size and quality for observation and interpretation of all modalities.

 

Comment: The display or DICOM header must identify the parent institution, the name of the patient, second patient identifier (such as MRN or DOB), the date and time of the study. The ECG must also be displayed.


                        1.1.1.6B        Instrument settings to enable optimization of ultrasound enhancing agents (UEAs).


                        1.1.1.7B        Instrument settings to enable optimization of tissue Doppler imaging.

 

1.1.1.8B        Range or depth markers must be available on all displays.

 

1.1.1.9B        Capabilities to measure the distance between two points, an area on a 2-D image, blood flow velocities, time intervals, and peak and mean gradients from spectral Doppler studies.

 

1.1.1.10B        At least two imaging transducers, one of low frequency (2-2.5 MHz) and one of high frequency (3.5 MHz or higher); or a multi-frequency transducer which includes a range of frequencies specific to the clinical needs in adult echo.

 

Comment: A transducer dedicated to the performance of non-imaging continuous wave Doppler must be available at each site.

 

1.1.1.11B        An audible output must be present at the time of acquisition.

 

1.1.1.12B     Machines with some, but not all of the above, equipment may be used for limited or directed echocardiographic examinations. However, machines utilized for complete diagnostic procedures must include all of the above listed capabilities.

 

STANDARD – Procedure Volumes

 

1.2B         The annual procedure volume must be sufficient to maintain proficiency in examination performance and interpretation.

 

(See Guidelines below for further recommendations.)

STANDARD – Indications, Ordering Process and Scheduling

 

1.3B         Transthoracic Echocardiography testing is performed for appropriate indications.1

 

1.3.1B         Verification of the Indication – A process must be in place in the facility for obtaining and recording the indication. Before a study is performed, the indication must be verified and any additional information needed to direct the examination must be obtained.

 

1.4B         Transthoracic echocardiography testing is appropriately ordered and scheduled. Appropriate Use Criteria (AUC) documents pertaining to echocardiography must be available for review in the facility.

 

1.4.1B         Ordering Process – The echocardiogram order and requisition must clearly indicate the type of study to be performed, the reason(s) for the study and the clinical question(s) to be answered. The signed (electronic or handwritten) order/requisition must be present in the medical record of the patient.

 

1.4.2B         Definition of Procedure Types and Protocols

 

1.4.2.1B        Complete Studies:

 

i.            A complete imaging study is one that examines all of the cardiac chambers and valves and the great vessels from multiple views, then uses the available information to completely define any recognized abnormalities.

ii.          A complete Doppler study is one that examines every cardiac valve, and the atrial and ventricular septa for antegrade and/or retrograde flow. In addition, a complete Doppler study provides functional hemodynamic data.

 

1.4.2.2B        Limited Study: A limited study is generally only performed when the patient has undergone a complete recent examination and there is no clinical reason to suspect any changes outside the specific area of interest. A limited study generally examines a single area of the heart or answers a single clinical question.

 

1.4.3B         Scheduling – Sufficient time must be allotted for each study according to the procedure type. The performance time allotted for a complete (imaging and Doppler) transthoracic examination is 45 to 60 minutes from patient encounter to departure. An additional 15 to 30 minutes may be required for complicated studies.

 

1.4.3.1B        An urgent study must be performed in the next available time period.

 

1.4.3.2B        A stat study must be performed as soon as possible, preempting routine studies.

 

1.4.3.3B        Availability for Emergencies: Qualified personnel and equipment must be available for urgent or stat studies outside normal working hours in inpatient facilities or where appropriate.

 

(See Guidelines below for further recommendations.)

 

STANDARD – Techniques

 

1.5B         Examination performance must include proper technique.

 

1.5.1B         All procedures must be explained to the patient and/or parents or guardian. The patient’s height, weight and blood pressure must be measured and recorded prior to or during the examination.

 

1.5.2B         Echocardiography examinations of the heart must examine all cardiac chambers and structures. The course and extent of disease must be documented.

 

1.5.3B         Elements of study performance include, but are not limited to:

 

1.5.3.1B        proper patient positioning;

 

1.5.3.2B        transducer selection and placement;

 

1.5.3.3B        optimization of equipment gain and display settings;

 

1.5.3.4B        utilization of appropriate Doppler technique (including proper Doppler alignment) and measurements;

 

1.5.3.5B        representative image storage of all images and data;

 

1.5.3.6B        timely report generation and communication of results; and

 

1.5.3.7B        performance of a 2-D/M-Mode/Doppler examination according to the facility specific and appropriate protocol that incorporate all views and imaging planes mandated by the Standards 1.5.1B, 1.5.2B.

 

1.5.4B         Elements of study quality include, but are not limited to:

 

1.5.4.1B        definition of endocardium;

 

1.5.4.2B        display of standard (on axis) imaging planes (e.g., avoidance of foreshortening);

 

1.5.4.3B        delineation of the details of valvular anatomy;

 

1.5.4.4B        measurements of left heart chamber dimensions and volumes (indexed when appropriate) from standard imaging planes;

 

1.5.4.5B        optimal recording and evaluation of Doppler flows (which are aligned to the Doppler beam and parallel to flow);

 

1.5.4.6B        accurate spectral Doppler recording and recording of abnormal Doppler flow signals in multiple views; and

 

1.5.4.7B        adherence to the facility specific protocol including sequence with allowances for additional views.

 

STANDARD – Components of the Transthoracic Echocardiogram

 

1.6B         Transthoracic echocardiograms must be comprehensive and include standard components.

 

1.6.1B         Components of the Examination – A protocol must be in place that defines the components of the standard examination. Indications for performance of a complete and/or limited examination must be included.

 

1.6.1.1B        Complete Examination: Includes standard views from multiple planes including views of all cardiac structures and selected extracardiac structures. These include, but are not limited to:

 

i.            left ventricle;

ii.          right ventricle;

iii.         left atrium;

iv.        right atrium;

v.          aortic valve;

vi.        pulmonic valve;

vii.       mitral valve;

viii.     tricuspid valve;

ix.        proximal ascending aorta;

x.          aortic arch;

xi.        inferior vena cava; and

xii.       pericardium.

 

1.6.1.2B        Complete Doppler Study: Includes spectral Doppler and/or color flow interrogation of all normal and abnormal flows within the heart including the valves, the great vessels and the atrial and ventricular septa.

 

1.6.1.3B        Limited Examination: A limited study is generally only performed when the patient has undergone a complete recent examination and there is no clinical reason to suspect any changes outside the specific area of interest. A limited study generally examines a single area of the heart or answers a single clinical question. 7

 

(See Guidelines below for further recommendations.)

 

1.6.2B         The complete exam must include (except where technically unobtainable), but not be limited to:

 

1.6.2.1B        The following standard 2-D views:

 

i.           parasternal long axis view;

ii.          right ventricular outflow tract view;

iii.          right ventricular inflow view ;

iv.         parasternal short axis views (at the level of the aortic valve, left ventricle at the basal, mid and apical levels);

v.          apical four-chamber view;

vi.         focused view of the right ventricle;

vii.        apical five-chamber view;

viii.       apical two-chamber view;

ix.          apical long axis view (three-chamber);

x.           subcostal four-chamber view;

xi.          subcostal short axis view (when indicated);

xii.         subcostal IVC/hepatic vein view; and

xiii.        suprasternal notch view.

 

1.6.2.2B        The following 2-D, 3-D or M-Mode measurements of the left heart:

 

i.           left ventricular internal dimension and/or volume at end-diastole;

ii.          left ventricular internal dimension and/or volume at end-systole;

iii.         left ventricular posterobasal free wall thickness at end-diastole;

iv.         ventricular septal thickness at end-diastole;

v.          left atrial dimension at end-systole or left atrial volume index; and

vi.         aorta at the level of the sinuses of Valsalva; measurements of sinotubular junction and mid-ascending, as clinically indicated;

vii.        left ventricular ejection fraction.


Comment: Quantitated ejection fraction by 3-D volumes or 2-D biplane method of discs is preferred over visual estimate.

 

1.6.2.3B        The following standard Doppler flow evaluations:

 

i.           four cardiac valves – forward flow spectra for each valve, and any regurgitation, shown in at least two imaging planes with color Doppler;

ii.          also use of non-imaging Doppler Transducer to assess stenotic valves, valvular regurgitation or whenever indicated;

iii.         tricuspid regurgitation spectrum must always be sought with CW Doppler from multiple views for estimation of systolic right ventricular pressure when tricuspid regurgitation is present;

iv.         atrial and ventricular septa – color Doppler screening for defects;

v.          left ventricular outflow tract velocity;

vi.          velocity-time integrals of the left ventricular outflow tract and the aortic valve, when clinically indicated;

vii.        hepatic and pulmonary vein flow spectra, when clinically indicated;

viii.       For aortic stenosis, the systolic velocity must be evaluated from multiple transducer positions (e.g., apical, suprasternal and right parasternal). This must include interrogation from multiple views with a dedicated non-imaging continuous wave Doppler transducer (at least one clear envelope must be obtained). 

ix.         Diastolic Function Evaluation – LV diastolic function must be evaluated through a combination of PW and tissue Doppler techniques.



1.6.2.4B        Use of Ultrasound Enhancing Agents (UEAs) for Suboptimal Image Quality – UEAs are indicated for use when two or more LV segments cannot be visualized adequately for the assessment of LV function (LVEF and regional wall motion assessment) and/or in settings in which the study indication requires accurate analysis of regional wall motion. [10][11]

 

i.           If UEAs are used, there must be a written policy for the use of UEAs. Although hypersensitivity reactions are rare, laboratories that use UEAs must have policies in place for emergent resuscitation of patients who may experience serious side effects.

ii.          Cardiopulmonary resuscitation personnel and equipment must be readily available prior to ultrasound enhancing agent administration.

                                       iii.          If a UEA is not able to be used, or if UEAs do not provide adequate visualization, a policy must be available for recommended alternative imaging.


Comment: Poor endocardial border definition is defined as the inability to detect two or more segments. 


(See Guidelines below for further recommendations.)

 

1.6.3B         Use of Contrast for Suboptimal Image Quality: Contrast is indicated for use when two contiguous segments are not visualized in any three of the apical views (poor endocardial border definition) as it provides greater accuracy in determining left ventricular function.

 

1.6.3.1B        If contrast is used, there must be a written policy for the use of contrast agents.

 

1.6.3.2B        If contrast is not able to be used there must be a policy for alternative imaging.

 

Comment: Poor endocardial border definition is defined as the inability to detect two or more contiguous segments in any three of the apical views. 

 

(See Guidelines below for further recommendations.)

 

 

Section 1B: Adult Transthoracic Echocardiography Testing
Guidelines

 

 

1.4.3B            Additional time should be allocated for technical staff to complete analysis and documentation for each study.

                       A routine study on an inpatient should be performed on the same working day as ordered, unless otherwise specified. Outpatient studies should be assigned priority as defined by the referring physician and/or the indication of the study.


1.4.3.2B         The facility should have a policy defining STAT echocardiogram indications.

 

1.6.1B            For all imaging protocols, if any required view or Doppler signal cannot be adequately obtained, it should be recorded and labeled in order to demonstrate that it was attempted.


1.6.2B            Strain imaging should be considered when clinically indicated. If strain is performed it should be reported. 3D imaging should be performed when clinically indicated, and should be reported.

1.6.2.1B        Additional views that should be considered include:

                            •    Parasternal short axis view of the aorta and branch pulmonary arteries 
                            •    Apical four-chamber view focused on atria
                            •    Apical two-chamber view focused on the left atrium

1.6.2.2B        Additional measurements that should be considered include:

                            •    Linear measurement of the right ventricle, when clinically indicated
                            •    Measurement and description of the response of the IVC to respiration, when clinically indicated
 

 

1.6.2.3B         Velocity-time intergrals of hepatic and pulmonary vein flow spectra, when clinically indicated.


1.6.2.4B         UEAs should be used in the presence of poor endocardial border definition for quantification of chamber dimensions, volumes, ejection fraction and assessment of regional wall motion.

UEAs should also be used to assess conditions such as hypertrophic cardiomyopathy or when left ventricular thrombus is suspected.

Allergy kits should be available and easily accessible in all areas where UEAs are in use and expiration dates should be checked on a regular basis in accordance with local institutional or lab policies.