The IAC Standards and
Guidelines |
Click here for a printer-friendly PDF of the
Cardiovascular Catheterization Standards |
2.1A Adequate
facilities must be provided for all operations of the facility so that patient
comfort, safety, dignity and privacy are ensured as well as staff comfort and safety.
Procedure areas must have sufficient space, be well maintained and clean. There
should be adequate space for the cardiovascular catheterization personnel to
freely access the patient and for all staff to maintain the sterile field.
Physical space requirements include, but are not limited to:
(See Guidelines
below for further recommendations.)
2.1.1A waiting, reception
and patient/staff bathrooms;
2.1.2A patient education,
consultation and examination areas;
2.1.3A readily accessible
handwashing/sanitation for staff;
2.1.4A performance of
pre-test/post-procedures within appropriate proximity of the procedure area
including adequate space for performing resuscitation in case of emergency;
2.1.5A emergency
cardiovascular surgical support must be immediately available in case of
life-threatening procedural complications;
Comment: Refer to Part B
for procedural requirements requiring emergency cardiovascular surgical support
and protocol for the transfer of patients to a tertiary facility in the event
of an emergency.
Comment: Cardiovascular catheterization procedures on pediatric patients, as
well as patients of any age with complex congenital heart defects, should only
be performed at centers with experienced cardiovascular surgical staff and the
proper equipment to provide back-up for emergencies.13
Comment: Centers performing pediatric
cardiovascular catheterization should have an on-site pediatric intensive care
unit and an on-site pediatric cardiac surgery program, in addition to a
pediatric cardiac anesthesia service. The pediatric cardiovascular
catheterization laboratory (PCCL) should have access to rescue ECMO, in
addition to standard resuscitation methods and technologies.6,20,32
Comment: Centers performing adult
congenital heart defect (ACHD) cardiovascular catheterization should have an
on-site cardiac intensive care unit with an ACHD consultation service, an
on-site ACHD surgical program, and a cardiac anesthesia service. The ACHD
catheterization laboratory should have access to rescue ECMO, in addition to
standard resuscitation methods and technologies.6,20,32
2.1.6A adequate space,
facility configuration and doorways for the emergency transport of patients
from patient care areas and for emergency exit of staff;
2.1.7A the following
procedure room type/area must comply with all Standards listed above (2.1.1A
through 2.1.6A) and have or meet, but not limited to, the following6:
2.1.7.1A Dedicated
Cardiovascular Catheterization Suites: Cardiovascular catheterization procedure
rooms that may additionally offer, intracardiac echocardiography (ICE), transesophageal
echocardiography (TEE), and use of robotics, which must provide for/include,
but not limited to6,13,14,15,16:
i. positive airflow;
ii. high flow oxygen
and vacuum for suctioning;
iii. medical gas
availability;
iv. substerile scrub
area;
v. patient
post-procedural care area(s);
vi. Room utilities:
Adequate utilities based upon the types of procedures and workload. These
utilities include water taps, lighting, electrical outlets, emergency power,
telephones, heating/cooling and ventilation.
vii. General room
lighting: Overhead and task lighting must be adequate to perform cardiovascular
catheterization procedures and for clinical evaluation and treatment of the
patient. The overhead lighting must be able to be dimmed during fluoroscopy. It
is recommended that the overhead lighting be controlled by a foot pedal used by
the operating physician.
· Additionally, the
procedure room must have surgical lighting.
viii. Room power: The
facility must have a plan that outlines the response to unexpected power loss
or computer function, such as movement of the patient to another similarly
capable procedure room in the immediate vicinity.
· When normal power
is not available, emergency power should provide a minimum of
10 minutes of fluoroscopy, and at least one hour of backup power for the
computers, monitoring equipment and ancillary equipment.
· For systems
ordered after July 2011, there should be sufficient emergency power supply to
run fluoroscopy for a duration of one hour and to run the remainder of the
x-ray system components including lighting, for a minimum of 24 hours.
· Utilization of
emergency power must be visible by the operator at the normal working position.
· An uninterruptible
power supply for all computer equipment is required.
· X-ray equipment
and computers should not require rebooting during transition between normal and
emergency power or during power line instabilities.
ix. cardiovascular
catheterization specific equipment:
· contrast
injectors.
x. defibrillator;
xi. electrocardiogram and
hemodynamic monitoring equipment capabilities as described in Standard 2.4.4A;
xii. radiolucent table
to include, but not limited to:
· height adjustable;
· support more than
159kg/350 lbs.;
· longitudinal and
lateral displacement; and
· length and width
appropriate to accommodate the patient population being treated (e.g.,
pediatric, adult, bariatric).
xiii. non-invasive blood
pressure monitor;
xiv. supplies specific
to the procedure(s) being performed;
xv. emergency
equipment and supplies as required by Standard 2.4.3A;
xvi. adequate space
must be provided to facilitate the use of emergency support equipment to
include, but not limited to:
· cardiopulmonary
bypass;
· extracorporeal
membrane oxygenation;
· intra-aortic
balloon pump;
· Impella;
· other.
xvii. a fixed
radiographic imaging system with flat-panel fluoroscopy offering cardiovascular
catheterization laboratory-quality imaging;
xviii. radiation shielded
barriers that meet state and federal requirements; and
xix. fluoroscopy
equipment must comply with requirements set by the Standards (refer to Appendix A).
Comment: A bi-plane unit is recommended
for procedures involving patients with congenital heart disease.
2.1.7.2A Combined Hybrid
Laboratories/Hybrid Surgical Suites: These are operating surgical rooms
offering cardiovascular catheterization procedures such as; valve
interventions, structural heart interventions, complex ACHD and pediatric
interventions, which must provide for/include, but not limited to6,20,21,22,23:
i. positive airflow;
ii. high flow oxygen
and vacuum for suctioning;
iii. medical gas
availability:
· medical air;
· nitrogen;
· nitrous oxide;
· oxygen.
iv. waste gas lines;
v. substerile scrub
area;
vi. patient
post-procedural care area(s);
vii. Room utilities:
Adequate utilities based upon the types of procedures and workload. These
utilities include water taps, lighting, electrical outlets, emergency power,
telephones, heating/cooling and ventilation.
viii. General room
lighting: Overhead and task lighting must be adequate to perform cardiovascular
catheterization procedures and for clinical evaluation and treatment of the
patient. The overhead lighting must be able to be dimmed during fluoroscopy. It
is recommended that the overhead lighting be controlled by a foot pedal used by
the operating physician.
· Additionally, the
procedure room must have surgical lighting.
ix. Room power: The
facility must have a plan that outlines the response to unexpected power loss
or computer function, such as movement of the patient to another similarly
capable procedure room in the immediate vicinity.
· When normal power
is not available, emergency power should provide a minimum of
10 minutes of fluoroscopy, and at least one hour of backup power for the
computers, monitoring equipment and ancillary equipment.
· For systems
ordered after July 2011, there should be sufficient emergency power supply to
run fluoroscopy for a duration of one hour and to run the remainder of the
x-ray system components including lighting, for a minimum of 24 hours.
· Utilization of
emergency power must be visible by the operator at the normal working position.
· An uninterruptible power supply for all computer equipment is required.
· X-ray equipment
and computers should not require rebooting during transition between normal and
emergency power or during power line instabilities.
x. cardiovascular
catheterization-specific equipment:
· contrast
injectors.
xi. defibrillator;
xii. electrocardiogram
and hemodynamic monitoring equipment capabilities as described in Standard
2.4.4A;
xiii. radiolucent table
to include, but not limited to:
· height adjustable;
· support more than
159kg/350 lbs.;
· longitudinal and
lateral displacement; and
· length and width
appropriate to accommodate the patient population being treated (e.g.,
pediatric, adult, bariatric).
xiv. non-invasive blood
pressure monitor;
xv. supplies specific
to the procedure(s) being performed;
xvi. emergency equipment
and supplies as required by Standard 2.4.3A;
xvii. snares and other
retrieval devices;
xviii. adequate space
must be provided to facilitate the use of emergency support equipment to
include, but not limited to:
· cardiopulmonary
bypass;
· extracorporeal
membrane oxygenation;
· intra-aortic
balloon pump;
· other.
xix. a fixed
radiographic imaging system with flat-panel fluoroscopy offering cardiovascular
catheterization laboratory-quality imaging;
xx. radiation shielded
barriers that meet state and federal requirements;
xxi. fluoroscopy
equipment must comply with requirements set by Standards (refer to Appendix A).
Comment: A bi-plane unit is recommended
for procedures involving patients with congenital heart disease.
(See Guidelines below for further
recommendations.)
2.1.7.3A Pediatric
Cardiovascular Catheterization Suites: Procedure rooms performing pediatric
cardiovascular catheterization procedures have similar requirements as that of
rooms detailed in Standards 2.1.7.2A and 2.1.7.3A with the exception of the
following requirements, which must include, but not limited to1,6:
i. pediatric
resuscitation equipment;
ii. pediatric
appropriate medication dosages;
iii. inventory of
pediatric catheters;
iv. inventory of
pediatric basic supplies;
v. fluoroscopy equipment
must comply with requirements set by the Standards (refer to Appendix A);
vi. fluoroscopy equipment must allow for digital acquisition or saved
fluoroscopy.
Comment: Centers performing pediatric
cardiovascular catheterization should have an on-site pediatric intensive care
unit and an on-site pediatric cardiac surgery program, in addition to a
pediatric cardiac anesthesia service. The pediatric cardiovascular
catheterization laboratory (PCCL) should have access to rescue ECMO, in
addition to standard resuscitation methods and technologies.6,20,32
Comment: Centers
performing adult congenital heart disease (ACHD) cardiovascular catheterization
should have an on-site cardiac intensive care unit with an ACHD consultation
service, an on-site ACHD surgical program, and a cardiac anesthesia service.
The ACHD catheterization laboratory should have access to rescue ECMO, in
addition to standard resuscitation methods and technologies.6,20,32
Comment: Facilities offering
cardiovascular catheterization procedures and personnel participating in these
programs must have a protocol for emergency response when the need arises.
There must be a mechanism in place to activate a rapid operating room response
team that is capable of performing emergency surgery. This “disaster
plan” should be regularly tested on a scheduled basis so that each member
of the team knows exactly what to do and how to accomplish their role. This
plan must be recorded as part of the written standard operating procedure of
every extraction laboratory or operating room.
(See
Guidelines below for further recommendations.)
2.1.8A The control
room/area must have or meet:
2.1.8.1A if the procedure
room is contiguous with the control room, a leaded wall with a large leaded
viewing window;
2.1.8.2A a full duplex
intercom system;
2.1.8.3A desk space
adequate to accommodate:
i. fluoroscopy
monitors; and
ii. a recording
system.
(See Guidelines below for further
recommendations.)
2.1.9A The substerile
entrance(s) must have or meet:
2.1.9.1A entrance with
dedicated or shared between adjacent procedure rooms;
2.1.9.2A entrance for
patient transport from the prep area to the cardiovascular catheterization
laboratory(s); and
2.1.9.3A egress that
connects to hallways leading to the hospital wards and other areas.
2.2A Adequate space
must be provided for the interpretation of examination results and preparation
of reports.
(See Guidelines below for further
recommendations.)
2.3A Adequate space
must be provided for:
2.3.1A the storage must
ensure confidentiality of data and be safe from fire/flood;
2.3.2A patient records,
reports and digital data storage areas;
2.3.3A administration
records and support areas; and
2.3.4A equipment/supply
storage areas.
(See
Guidelines below for further recommendations.)
2.4A Equipment and
instrumentation used in the performance of cardiovascular catheterization
procedures must be appropriate, in good working condition and should substantially
comply with the International Electrotechnical Commission (IEC) 60601-2-43
interventional standard.
2.4.1A All equipment and
instrumentation must be routinely inspected for safety and proper functionality
and records of the inspections kept on file. Equipment and instrumentation must
include at a minimum the following17,18,45:
2.4.1.1A a fixed
(recommended) or portable, single or biplane (recommended) cine angiography
system that must meet the following specifications;
2.4.1.2A digital subtracted
cine angiography imaging capable of cine frame rates from 15 to 60 frames per
second;
2.4.1.3A when applicable,
digital subtraction angiography imaging of at least three frames per second,
with the ability to do program change of rate during the image acquisition run;
2.4.1.4A when applicable,
high quality, subtracted digital radiographs;
2.4.1.5A road-mapping with
ability to refer back to an unsubtracted live image;
2.4.1.8A dose measurement
capability;
2.4.1.9A Digital Imaging
and Communications in Medicine (DICOM) compatible digital image storage with
capability of storing uncompressed images on portable format without loss of
image resolution;
2.4.1.10A ability to display
and review prior relevant images during the procedure;
2.4.1.11A minimum 9-inch
field of view (FoV) imaging intensifier at a Source Image Receptor Distance
(SID) of 30 inches;
2.4.1.12A minimum spatial
resolution of matrix of 512 x 512;
2.4.1.13A for pediatric
procedures, multiple focal spot sizes to accommodate the patient population
imaged at the facility;
2.4.1.14A image monitor
performance using the Society of Motion Picture and Television Engineers
(SMPTE) pattern;
2.4.1.15A for equipment
installed before 2006 that does not display cumulative dose and or dose area
product (DAP), documentation of fluoroscopy time and the number of images per
procedure is acceptable; and
2.4.1.16A radiation use must
be consistent with the “as low as reasonably achievable” principle
or ALARA radiation safety guidelines.
(See
Guidelines below for further recommendations.)
2.4.2A All equipment and
instrumentation must be routinely inspected for safety and proper functionality
and records of the inspections kept on file.
2.4.3A Emergency
equipment and supplies (response cart or kit) must include, but not limited to:
2.4.3.2A biphasic external
defibrillator with a backup defibrillator immediately accessible;
2.4.3.3A external pacemaker
equipment;
2.4.3.4A when applicable,
intra-aortic balloon pump (IABP);
2.4.3.5A when applicable,
cardiopulmonary bypass (CPB);
2.4.3.6A when applicable,
extracorporeal membrane oxygenation (ECMO);
2.4.3.7A emergency trays
for pericardiocentesis;
Comment: For interventional procedures,
there must be a process in place for immediate access to thoracotomy and
sternotomy trays.
2.4.3.8A standard Advance
Cardiac Life Support (ACLS) medications (including a master list with
verification of expiration date) to include, but not limited to:
i. epinephrine;
ii. atropine;
iii. dopamine;
iv. vasopressin;
v. adenosine;
vi. amiodarone;
vii. magnesium sulfate;
viii. calcium chloride;
ix. potassium
chloride;
x. sodium
bicarbonate; and
xi. sedative reversal
agents:
· flumazenil;
· naloxone.
2.4.3.9A endotracheal
intubation equipment to include:
i. sedative(s);
ii. paralytic
agent(s); and
iii. anesthetic agent.
2.4.3.10A resuscitator bag
and mask;
2.4.3.11A non-rebreather
mask;
2.4.3.12A arterial blood gas
kits;
2.4.3.13A separate
monitoring system for ECG and hemodynamics to include:
i. pressure
transducer; and
ii. end-tidal carbon
dioxide monitor.
2.4.4A Monitoring
equipment must be available to perform:
2.4.4.1A intravascular
pressure;
2.4.4.2A non-invasive blood
pressure;
2.4.4.4A activated
coagulation time (ACT) analyzer;
2.4.4.5A electrocardiogram
(ECG);
i. 12-lead surface
ECG.
2.4.4.6A capnography (CO2)
monitoring is recommended for use with moderate sedation.
2.4.5A Procedure table(s)
must be radiolucent, motorized and have the following capabilities:
2.4.5.2A support more than
159kg/350 lbs.;
2.4.5.3A longitudinal and
lateral displacement; and
2.4.5.4A length and width
appropriate to accommodate the patient population being treated (e.g.,
pediatric, adult, bariatric).
2.4.6A Additional systems
and applications may be used during the course of performing a cardiovascular
catheterization procedure:
2.4.6.1A Intracardiac
echocardiography (ICE) systems using a linear phased array transducer that
produces a 90-degree image longitudinal to the catheter and/or a rotational
transducer to display a 360-degree image perpendicular to the catheter.
(See
Guidelines below for further recommendations.)
2.4.6.2A robotic navigation
systems; and
2.4.6.3A transthoracic
echocardiography (TTE) and/or transesophageal echocardiography (TEE) (refer
to 2.4.7A for additional requirements).
.
(See Guidelines below for further recommendations.)
2.4.7A Ultrasound imaging
equipment:
2.4.7.1A Must meet the IAC
Standards and Guidelines for Adult and/or Pediatric Echocardiography
Accreditation for equipment and must include:
i. color flow
Doppler;
ii. imaging
frequencies appropriate for the structures evaluated;
iii. Doppler
frequencies appropriate for the vessel evaluated;
iv. a visual display
and capability of permanent recording of the image; and
v. a visual display,
audible output and capability for a permanent recording of the Doppler waveform
and corresponding image.
2.4.8A Adequate
disposable supplies must be immediately available. These include U.S. Food and Drug
Administration (FDA) approved catheters, wires, stents, balloons and embolic
protection devices. Non-FDA approved devices may also be used as permitted by
law.
2.4.9A Ancillary
equipment as appropriate (e.g., monitoring equipment, blood coagulation testing
equipment, workstations, picture archiving communication system [PACS],
radiation protection for personnel [aprons and thyroid shields, portable shield
either on wheels or suspended from ceiling]).
2.5A There must be a
comprehensive Quality Improvement (QI) program to provide a standard of
measurement for system performance and the documentation of any variance
thereof.
2.5.1A A QI Committee
should be appointed to provide oversight to the equipment and instrumentation
quality control (QC).
2.6A Fluoroscopic
system QC testing must include a comprehensive evaluation of the system
components, image performance and radiation output limits as outlined in the
FDA Code of Federal Regulations (CFR) Title 21 subchapter J, Parts 1010 and
1020 and applicable FDA guidance documents.11
2.7A Image quality
requirements, radiation output limits, and other fluoroscopic performance
requirements must also comply with the health-code regulations of the state in
which the facility is located.
2.8A The site-appointed
qualified medical physicist must complete the performance evaluations at
equipment installation and annually, unless state regulations require more
frequent testing. Equipment performance evaluations are recommended
semi-annually to include radiation output measurements, system quality control
tests and image quality performance measurements.
2.9A The site-appointed
medical physicist must perform a radiation safety survey to ensure that
occupational workers and members of the public are shielded according to state
regulation. This must be performed prior to installation of each new
angiographic imaging system. A documented radiation safety survey of the
cardiovascular catheterization suite and adjacent areas that has been accepted
by the State Radiation Program fulfills this requirement.
2.10A A radiation safety
survey must be performed on all renovated or newly constructed interventional/
cardiovascular catheterization suites and adjacent areas. This must be
performed prior to first patient use. This survey must confirm that the levels
of radiation protection are in conformance with the State Radiation Program.
2.11A All spaces outside
the procedure rooms should provide adequate protection for full time occupancy
by non-radiation workers. This recommendation includes the control room.
2.12A Preventive
maintenance (PM) on all cardiovascular catheterization equipment listed in
Standards 2.4.4.A, 2.4.5A, 2.4.6A, 2.4.7A and 2.4.8A is required according to
the manufacturer’s recommendations.
2.13A Preventive
maintenance (PM) service is recommended periodically per the manufacturers’
recommendations for each angiographic system at the facility.
2.14A Ancillary
equipment (e.g., hemodynamic monitoring equipment, blood gas and coagulation
testing equipment, workstations, PACS, lead aprons, suction, oxygen lines, etc.)
should also be included in a PM program.
2.15A The emergency
response cart or kit must be checked at least monthly, with documentation to
assure that all expected items are present and the supplies are not expired.
2.16A There must be a
process to check inventory of disposable supplies (e.g., catheters, wires,
balloons, stents, embolic protection devices, contrast, and portable oxygen
tank) on a regular basis to assure that these supplies are readily available
during a procedure.
2.17A All QC results
must be documented and reviewed.
2.17.1A Documentation of
the physicists’ evaluation, preventive maintenance and quality control
tests performed, and service records for all angiographic systems and ancillary
equipment must be maintained at the facility and available for review. The
reports must be signed and dated by the person(s) performing the tests.
2.17.2A Results of all QC
tests must be documented, archived and stored on film, in digital format, or on
other suitable media according to state requirements (if applicable).
2.1A
The participation of an ergonomics expert in the planning should be considered
as a measure to comply with Occupational Safety and Health Administration
standards.
2.1.7A
The Guidelines for Design and Construction of Hospitals and Health Care
Facilities published by the American Institute of Architects and the Facility
Guidelines Institute provide space and functionality standards for
cardiovascular catheterization laboratories with a goal to improve work flow
in the cardiovascular catheterization environment.
The minimal procedural area of a complete
cardiovascular catheterization laboratory (not including control room space) is
350 square feet of clear floor area.
There should be a minimum of 8 feet of
clear space between the wall and the edges of each side of the patient table
when it is positioned at the isocenter.
Enough clearance at the head of the bed
should be allocated for anesthesia equipment on either side and sterile access
to jugular vein entry sites, if employed, while allowing for free range of
movement of the fluoroscopy C-arm.
Current electrical system regulations for
health care facilities should follow Article 517 of the National Electrical
Code (NEC) Handbook
The air flow/heating, ventilation,
and air conditioning design should comply with the Guidelines for Environmental
Infection Control in Health-Care Facilities Recommendations of the Centers for
Disease Control and Prevention and the Healthcare Infection Control Practices
Advisory Committee document.
Lighting should include an overhead light
on an articulating arm, 2 x 2 feet lighting squares to flood the main
procedure area, and a dedicated workspace light for the nursing/anesthesia
area.
2.1.7A and
2.1.8A The ideal sound/communication
system is an always-on, full-duplex, two-way intercom system.
Network cabling and hardware should have a
minimum capability of support for gigabit Ethernet speed.
2.1.7.2A Hybrid
rooms should be in close proximity to operating room(s) or catheterization
suite(s) and located on a clear core or semirestricted corridor where scrubs,
hats and masks are required.6
2.1.8.3A An
additional 45 inches of desk space is suggested for a two-monitor reading
station or single-monitor workstation.
2.2A and
2.3A Electronic storage of cardiovascular
catheterization data should be Health Insurance Portability
and Accountability
Act (HIPAA) compliant. Data should be maintained for at least the minimum
duration as
determined by each state.
2.4A
Integrated data display systems provide flexibility and efficiency
in data display; it is advisable to have separate backup monitors in case of
failure.
2.4.1A
It is important to achieve the lowest possible noise signal with all recording
systems.
2.4.6.1A
Intracardiac Echocardiography (ICE) may be useful as an adjunctive imaging
modality during complex procedures.
2.4.6.3A
Transthoracic echocardiography and transesophageal echocardiography should be
readily available for emergency use and for adjunctive imaging in selected
cases.